Drager Polytron TX

Code: Polytron TX
  • Obsolete
  • Toxic
  • Replacements available

In StockMake An Enquiry

Drager Polytron TX

This product is now obsolete. The equivalent is the Polytron 5100.

Polytron TX was a permanent gas monitor for detecting toxic gases or O2. Zoned area approvals ATEX, UL and CSA.

Two mounting holes were incorporated into the Powder-coated aluminium transmitter housing for installation.

Toxic gases and Oxygen are configured, and the user could choose the ranges below.

CO 0 to 100 / 300 / 500 ppm
H2S 0 to 20 / 50 / 100 ppm
O2 0 to 25 Vol%
Cl2 0 to 5 ppm *)
H2 0 to 500 ppm *)
NH3 0 to 100 / 300 / 500 ppm *)
NO2 0 to 20 ppm *)
SO2 0 to 20 ppm *)

Output to preventive devices, a building management system, or a control room is was provided by a 4-20mA signal—fault signal indicated by 2mA or less.  Maintenance 3, 4 or 5mA.

Electrical supply 16 to 30 V DC, 2-wire

For commissioning and calibration, tapping the magnet wand on the face will allow external access to the calibration and menus of the detector.

Part Numbers

4543405 Polytron TX without sensor; UL, CSA, ATEX, suitable for CO, H2S, O2
4543435 Polytron TX without sensor; general purpose, suitable for Cl2, H2, NH3, NO2, SO2
4543428 Magnetic wand for menu access

Sensors

6809605 CO
6809610 H2S
6809630 O2
6809665 Cl2
6809685 H2
6809680 NH3 LC
6809645 NH3 HC
6809655 NO2
6809660 SO2

Enquire About This Product

Title Size
PolytronTX 177 KB

Additional information

Weight 0.09 kg