Geotech G200 N2O Analyser

Code: G200-00N
  • Leak detection
  • User-settable alarms
  • Data download for graphing & reporting

In StockMake An Enquiry

The G200 N2O analyser is designed to safety check background and breathing zone levels of N2O (0-1000ppm) in medical applications.

Features

  • 0 – 1000 ppm N2O.
  • Storage for 1000 readings.
  • TWA calculated.
  • EH40 occupational exposure limits calculated.

Benefits

  • Accurate verification of exposure limit breaches.
  • Dual-purpose background analyser or personal analyser.
  • Leak detection for N2O storage.
  • Highly portable is running efficiently.

Applications

  • Operating theatres.
  • Dental practices.
  • Veterinary clinics.
  • X-ray departments.

Brand

  • Geotech/QED Environmental Systems.

Order Information

Analysers:

G200-00N G200 Analyser, N2O 0-1000 PPM
G200-00L G200 Analyser, N2O 0-1000 PPM (US version)

Accessories:

USBLEAD2 USB Lead
068296/S Spare Sample Filters (pack of 5)
G1.6 Sample Tube Kit
050227 Hard Carry Case
CDA7.6 Spare Calibration Gas (Various gas options available)
2008248/S Mains Charger including Worldwide Adaptors
G1.11 Soft Carry Case
G1.10 Soda Lime Filter Kit
G1.4 Analyser Data Manager Software
G1.1 Regulator and Tubing for Calibration Gas

Old part numbers:

G200-00B G200 Analyser – no longer available for purchase
G200-00P G200 Analyser – no longer available for purchase

    Enquire About This Product


    Title Size
    G200 Datasheet 1 MB
    G200 Manual 2 MB

    Additional information

    Weight 0.09 kg

    Specifications

    POWER SUPPLY
    Battery type Li Ion
    Battery life 12 hours (10 hours with pump)
    Battery lifetime 600 cycles
    Battery charger 5Vdc external power supply and internal charging circuit
    Charge time Approximately 4 hours from complete discharge
    Alternative power 5Vdc power supply
    GAS RANGES
    Gases measured N2O – By custom dual wavelength infra-red cell with reference channel
    Range N2O – 0 – 1,000ppm
    N2O – 0- 10,000ppm (leak detection mode, indication only)
    Typical accuracy* N2O Resolution: 1ppm
    ± 5ppm for 0- 100ppm after calibration
    ± 1.5% of range for 101ppm and above, after calibration
    Response time, T90 N2O ≤ 40 seconds
    *Typical accuracies All typical accuracies quoted are after calibration plus accuracy of calibration gas used.
    FACILITIES
    Visual and audible alarm User selectable N2O and TWA alarms
    Communications USB type B mini-connector, HID device class
    Data Storage 1000 reading sets + 270 events
    50 site ID’s and 300 sample point ID’s
    PUMP
    Flow 100cc / min typically
    ENVIRONMENTAL CONDITIONS
    Operating temperature 0°C to 50°C
    Barometric pressure 500 to 1500mb
    Relative humidity 5% to 95% non condensing
    IP rating IP40
    PHYSICAL
    Weight 500 grams
    Size L 165mm, W 100mm, D 55mm
    Case material ABS / Polypropylene with Silicone Rubber Inserts
    Keys 17 Resin capped Silicone rubber keys
    Display Liquid crystal display, 128 x 64 pixel
    With RGB LED back-light
    Gas Sample Filters User replaceable PTFE water trap filter, G1.10 – Soda lime CO2 filter
    CERTIFICATION
    EN 50270 : 2006 Electromagnetic compatibility- Electrical apparatus for the detection and measurement of combustible gases, toxic gases or oxygen
    EN 61010-1 : 2010 Safety requirements for electrical equipment for measurement, control, and laboratory use. Part 1: General requirements

      Enquire About This Product