TSI mass flow meters are designed for Laboratory, Industry or Medical applications demanding high accuracy and a choice of data output options. The sensor uses proprietary platinum film to measure gas flow.
There are nine TSI mass flow meters, falling into two types; the high flow 4000 range, 0-300 L/min and the low flow 4100 series 0.01-20 L/min. All models are calibrated for air, additional calibrations for Oxygen and Nitrogen are available for both types. Low flow models can have Nitrous Oxide calibrations.
Two outputs are included analogue 0-10 VDC flow only, zero and span adjustable via RS232 and digital RS232 output.
Accuracy is ±2% at full-scale flow rate reading. If a full scale is 300 L/min, then the uncertainty for all readings is ±6 L/min. One TSI flow meter covers the same range as three or more “percent of full scale” devices, with better accuracy at all points. The rapid four-millisecond response gives accuracy when the flow is variable. This fast response time is ideal for closed-loop control systems and integrated volume measurements. Pressure and temperature measurements are also fast. A low-pressure drop minimises flow circuit back pressure and its impact on the system under test.
Features
- Turn-down ratios greater than 1000:1
- High Accuracy- pressure ±1 kPa & temperature ±1°C at flow greater than 1 Std L/min
- Digital and Analog outputs
- Multi-gas, user-selectable
- Fast 4 ms response flow changes
- Little pressure loss
- Measure flow rate, temperature, and pressure (display unit only) in one instrument
- Calibration certificate is NIST-traceable and included with purchase
Models – Sector Suitability Detailed
40401 measures air, connection tube size 22mm ISO, Range L/min 0-300, medical, r&d and industry.
4040 measures air, O2, & N2 connection tube size 22mm ISO, Range L/min 0-300, medical, r&d and industry.
4043 measures air, O2, & N2 connection tube size 1/2″, Range L/min 0-200, industry.
4045 measures air, O2, & N2 connection tube size 3/4″, Range L/min 0-300, industry.
41401 measures air, connection tube size 1/4″, Range L/min 0.01-20, industry.
4140 measures air, O2, & N2 connection tube size 1/4″, Range L/min 0.01-20, medical, r&d and industry.
4143 measures air, O2, & N2 connection tube size 3/8″ Range L/min 0.01-20, medical, r&d and industry.
41403 measures air, O2, & N2O connection tube size 1/4″ Range L/min 0.01-20, medical.
41433 measures air, O2, N2 & N2O connection tube size 3/8″ Range L/min 0.01-20, medical.