Honeywell ACM 150 FTIR Analyser

Code: 76-ACM150-115-1G2
  • Gas cell path: 5.0 m path length.
  • Scanning rate: ≤ 15 sec per point.
  • Analysis method: Continuous scan FTIR (Fourier Transform Infrared) analyser.

In StockMake An Enquiry

Lower your cost-per-point gas monitoring with next-generation FT-IR technology.

The ACM 150 uses versatile and proven Fourier Transform Infrared Technology to detect a wide range of volatile chemicals. It uses the infrared fingerprint of each gas for identification, which is free of cross-interferences. It can also identify unknown chemicals or odours.

Detection of new gases can be added at any time. By choosing between two sample cells, the ACM can detect either ppm levels for toxic gases or per cent levels for flammable gases.

The gas selections and alarm configuration for each point can be easily configured or changed either from the front panel or remotely via a web browser. Up to 14 gases can be detected at each point. Up to 4 points can be sampled together for increased speed. Because of its modular design and built-in diagnostics, the ACM is easy to maintain.

An optional LonWorks interface is ideal for connection to Honeywell Analytics’ Life Safety System.

Integrate gas detection into a single unit:

  • Detect multiple gases, chemicals, or odours in an industrial process*.
  • Overlay the spectrum of detected gas against a library of spectra for precise detection as a user validation tool.
  • Up to 40 measuring points can be detected sequentially.
  • 80 configurable relay points for increased flexibility.

Promote uptime:

  • Sealed aluminium body prevents moisture intrusion, protecting critical internal components.
  • Double IR scan ensures reliable performance and reduces false alarms.
  • Optimised built-in measurement and self-diagnostic capabilities eliminate the need for on-site gas calibration.
  • Desiccant replacement cycle of 3 to 5 years.

Boost gas monitoring performance:

  • Composite sampling of up to four measuring points provides faster measurement.
  • Easy-to-service gas cell with stainless-steel body construction prevents costly downtime.
  • Better signal-to-noise ratio (SNR) performance allows for lower LDL measurements.

Simplified touchscreen operation:

  • Improved HMI with a modernised solid-state computer operating on a world-proven operating system.
  • Large user-friendly touchscreen interface.
  • Flexible communications options meet unique site requirements.
  • Robust solid-state drive.
  • Embedded web server for remote access.
  • Refer to the ACM 150G2/100 Gas List or ask your regional sales rep for more information. More than two hundred gases are detectable. Our engineers may be able to generate a new method for gases not listed.

Order Information

ACM 150 Base Unit
76-ACM150-115-1G2 ACM150 120VAC, primary venturi pump version
76-ACM150-115-2G2 ACM150,120VAC, primary electric pump version
76-ACM150-230-1G2 ACM150,240VAC, primary venturi pump version
76-ACM150-230-2G2 ACM150,240VAC, primary electric pump version
ACM Method Development
100-12899-00 Method Generation / Setup (per gas)
Chemistry / Application Support & Development
1991-0199 One hour of laboratory/development time (4 hours minimum)
1991-0200 Full day of laboratory/development time (8 hours)
1991-0202 Full week of laboratory/development time (40 hours)
ACM 150 Options and Accessories
76-150-14393-04-1 10-point Sample Manifold includes a Dust Filter. Up to 3 max may be added to the base unit – 40 points total.
76-165-14594-00-1 Kit, Line Leak,3/8T Check Valve. Note: 1 per sample point required.
RSD0013 Set up of method
0235-0252 FEP tubing 0.25″ ID x 0.375″ OD, 300m length
76-215-12477-00-B ACM internal hardware for the Line Leak Option for dual electric pump systems.
76-215-12475-00-D ACM internal hardware for the Line Leak Option for Venturi pump systems.
ACM 150 Serial Communications Interfaces
76-150-14392-02 Kit, Data Interface, Control Net B3514
76-150-14398-02 Kit, Data Interface, Control Net, Redundant B3514
76-150-14391-10 Kit, Data Interface, Modbus TCP/IP
76-150-14390-04 Kit, Data Interface, Profibus DP
ACM 150 Spares
76-465-15101-00 PKI bench w/plate
76-441-15551-00 Gas cell for PKI bench
76-150-15300-00-C Network switch, Cisco, including Bracket
76-150-14500-01-A Kit DVD /CD-Rom drive inc Mtg brkts
76-150-14500-00-B DVD/CD-ROM drive with mounting brackets
76-150-14600-00-A Kit, Single Board Computer ARK1123
3014B1212 ACM100 complete upgrade kit-(A Spectrum two/Gas cell and a bench mounting kit, an HMI PC and PC Bracket, a touch monitor, a router and an SBC)
3014B1213 ACM150 HMI upgrade kit (A HMI PC, a PC bracket, and a touch monitor)
3014B1214 A spare part, Touch LCD module for ACM150 G2
3014B1215 A spare part, HMI PC module for ACM150 G2
ACM 100 Upgrade Kits
76-165-14582-00-1 Kit, Dust Filter 3/8
ACM 100 Spares
76-215-12477-00-B ACM internal hardware for the Line Leak Option for dual electric pump systems. Note: Assy, Line Leak requires qty 1 off item 76-220-13588-00-A when used for ACM 100.
76-215-12475-00-D ACM internal hardware for the Line Leak Option for Venturi pump systems. Note: Assy, Line Leak requires qty 2 off item 76-220-13588-00-A when used for ACM 100.
76-220-13588-00-A Assy, Adaptor Solenoid Valve for new plug
76-120-15050-00 Software ACM 100 Archiving Tool

    Enquire About This Product


    Title Size
    ACM150 Datasheet 3 MB

    Specifications

    Performance Specifications
    Analysis Method Continuous scan FT-IR analyzer
    (FT-IR = Fourier transform infrared) Gas Cell Path 5.0 m path length
    Gases Monitored Organic, PFC, CFC, HFC, Metal
    Organic, NF3 and a wide variety of other inorganic gases
    Lower Detection Limit 0.1 to 2.0 ppm for 5.0 m path cell, gas dependent
    Scanning Rate 15 sec per point
    Sample Point Capacity Available in 10, 20, 30, and 40 point configurations composite sampling up to 4 points
    Data Archiving 30 days for all spectra, 90 days for alarms
    Applicable Standards EN 50270, EN 61010, UL 61010
    Installation Indoor only
    Altitude Up to 2,000 m (6,562 ft.)
    Temperature range 5°C to 30°C (41°F to 86°F)
    Relative humidity 80% at temperatures up to 31°C (88°F), decreasing linearly to 50% RH at 40°C (104°F)
    Main supply voltage ±10% of the nominal voltage
    Overvoltage Category II
    Pollution degree 2
    Ingress protection IP 20
    Facilities Requirements
    Sample Line Tubing 1/4” ID x 3/8” OD polypropylene
    Sample Line Inlets 3/8” Swagelok® connector
    Electrical Supply 230 VAC, 10 A or 115 VAC, 20 A
    Purge Gas, N2 5 – 10 psi, 10 l/min
    Venturi Requirements 90 – 120 psi, 320 l/min
    Exhaust Pipe 1.5” NPT
    Overall Dimensions 66 x 34 x 25 inches (HxWxD) 1,676 x 864 x 635 mm (HxWxD)
    Weight
    Operating 349 kgs / 770 lbs
    Shipping 470 kgs / 1036 lbs
    Standard Outputs and Communications
    Relay Outputs contacts 80 x DPDT (double-pole double-thow), programmable by sample point and gas type
    Max. ratings 30 VDC, 2 A
    Interfaces Web based user interface (e-Diagnostics) Ethernet: HTML viewing / control; ODBC access Touchscreen user friendly interface (point and click)
    Optional Outputs and Communications via Gateways
    Interface LonWorks®, Allen Bradley®, Modbus®, Profibus®, OPC drivers others on request
    Options and Accessories
    Automatic sample line leak test User settable

      Enquire About This Product