Carbolite Gero HTRV-A High-Temperature Vertical Split Tube Furnaces
A single zone vertical split tube furnace range with a choice of maximum temperatures; 1600°C or 1700°C. Models available accommodate 70mm up to 100mm overall diameter work tubes.
The furnace and control box are separated by three metres of cable, complete with plug and socket. No stand is supplied so that a customer can integrate the furnace with their equipment. An L stand option is available for furnace support.
Split heating module allows either easy positioning of the work tube or positioning around reactors which have fixed end flanges. The split design may also allow faster cooling of samples. The control thermocouple is fitted in the centre of the heating zone. Cooling channels are engineered into the housing to aid with convection cooling of the outer case. The two furnace chamber halves consist of high-grade insulation plates with vertically hanging MoSi2 heating elements. A safety switch protects the operator by switching off the heating elements once the furnace is opened.
Standard features
- 1600° & 1700 °C maximum operating temperatures
- EPC3016P1 programmable temperature controller with 24 segments
- Over-temperature controller with independent thermocouple
- Vertical Orientation
- Accepts work tubes with outer diameters up to 100 mm for use with modified atmosphere
- Single-zone heated lengths of 120, 250, 500 or 700 mm
- Split design allows work tubes or reactors with fixed flanges to be accommodated
- High-grade type B thermocouple
- Low thermal mass ceramic fibre insulation
- Vertically hanging high-quality MoSi2 heating elements
- Ethernet communications
Options
- A range of sophisticated digital controllers, multisegment programmers, and data loggers with digital communication options
- A range of additional work tubes end seals and work tube packages for use with modified atmosphere and/or vacuum
- Longer heated lengths
- Stand for convenient and safe usage
- Vacuum packages with a choice of rotary vane pump or turbomolecular pump
- A wide selection of tube diameters and materials
- Insulation plugs and radiation shields to prevent heat loss and improve uniformity
- Oxygen sensor for inert gas packages
- 6 m length of cable between furnace body and control box with plug and socket
- Gas packages with manual or automatic valve for up to 3 gases
Model | Max temp °C | Furnace Ø mm | Heated lengths mm | Uniform length ± 5 °C mm |
Max Power W |
Single Zone Vertical Split Tube Furnaces | |||||
1600°C | |||||
HTRV-A 16/70/120 | 1600 | 70 | 120 | 50 | 4800 |
HTRV-A 16/70/250 | 1600 | 70 | 250 | 125 | 6000 |
HTRV-A 16/70/500 | 1600 | 70 | 500 | 250 | 13000 |
HTRV-A 16/70/700 | 1600 | 70 | 700 | 350 | 19000 |
HTRV-A 16/100/500 | 1600 | 100 | 500 | 13000 | |
HTRV-A 16/100/700 | 1600 | 100 | 700 | 19000 | |
1700°C | |||||
HTRV-A 17/70/250 | 1700 | 70 | 250 | 125 | 6000 |
HTRV-A 17/70/500 | 1700 | 70 | 500 | 250 | 13000 |
HTRV-A 17/70/700 | 1700 | 70 | 700 | 350 | 19000 |
HTRV-A 17/100/500 | 1700 | 100 | 500 | 13000 | |
HTRV-A 17/100/700 | 1700 | 100 | 700 | 19000 |